Ann Clin Biochem
November 2017
Background Interference from opiates in the Microgenics CEDIA® Buprenorphine assay is known to produce false-positive buprenorphine screening immunoassay results necessitating confirmatory buprenorphine testing by chromatography/mass spectrometry methods. Method We reviewed data on falsely positive buprenorphine immunoassay screen (cut-off ≥ 5 µg/L) but negative for buprenorphine by gas chromatography mass spectrometry (cut-off ≥ 5 µg/L) and had a positive opiate immunoassay result (cut-off ≥ 300 µg/L). The results were collected over three months, and the data were evaluated to determine whether there is an opiate immunoassay screen concentration below which a false-positive buprenorphine result will not occur.
View Article and Find Full Text PDF