In this work, a simple and efficient surfactant assisted pulsed two-phase electromembrane extraction (SA-PEME) procedure combined with gas chromatography (GC) has been developed for the determination of alfentanil, sufentanil and methadone in various samples. It has been found that the addition of anionic surfactant causes the accumulation of the cationic analytes at the SLM/solution interface resulting in an easier transfer of the analytes into the organic phase. The method was accomplished with 1-octanol as the acceptor phase and supported liquid membrane (SLM) by means of an 80 V pulsed electrical driving force and the extraction time of 20 min.
View Article and Find Full Text PDFIn order to understand the limitations of electromebrane extraction procedure better, a simple equivalent circuit has been proposed for a supported liquid membrane consisting of a resistor and a low leakage capacitor in series. To verify the equivalent circuit, it was subjected to a simulated periodical polarity changing potential and the resulting time variation of the current was compared with that of a real electromembrane extraction system. The results showed a good agreement between the simulated current patterns and those of the real ones.