An instrumented palpation sensor, designed for measuring the dynamic modulus of tissue in vivo, has been developed and trialled on ex vivo whole prostate glands. The sensor consists of a flexible membrane sensor/actuator with an embedded strain gauge and is actuated using a dynamically varying airflow at frequencies of 1 and 5 Hz. The device was calibrated using an indentation stiffness measurement rig and gelatine samples with a range of static modulus similar to that reported in the literature for prostate tissue.

View Article and Find Full Text PDF

Introduction: Minimally invasive radical prostatectomy (RP) (robotic and laparoscopic), have brought improvements in the outcomes of RP due to improved views and increased degrees of freedom of surgical devices. Robotic and laparoscopic surgeries do not incorporate haptic feedback, which may result in complications secondary to inadequate tissue dissection (causing positive surgical margins, rhabdosphincter damage, etc). We developed a micro-engineered device (6 mm2 sized) [E-finger]) capable of quantitative elasticity assessment, with amplitude ratio, mean ratio and phase lag representing this.

View Article and Find Full Text PDF

To systematically review the range of methods available for assessing elasticity in the prostate and to examine its use as a biomarker for prostate cancer. A systematic review of the electronic database PubMed was performed up to December 2012. All relevant studies assessing the use of elasticity as a biomarker for prostate cancer were included except those not studying human prostates or reporting a sensitivity, specificity or quantitative elasticity value.