We report the development of a method coupling microfluidics and a miniature mass spectrometer, applied to quantitation of drugs of abuse in urine. A custom digital microfluidic system was designed to deliver droplets of solvent to dried urine samples and then transport extracted analytes to an array of nanoelectrospray emitters for analysis. Tandem mass spectrometry (MS/MS) detection was performed using a fully autonomous 25 kg instrument.

View Article and Find Full Text PDF

A major design objective of portable mass spectrometers is the ability to perform in situ chemical analysis on target samples in their native states in the undisturbed environment. The miniature instrument described here is fully contained in a wearable backpack (10 kg) with a geometry-independent low-temperature plasma (LTP) ion source integrated into a hand-held head unit (2 kg) to allow direct surface sampling and analysis. Detection of chemical warfare agent (CWA) simulants, illicit drugs, and explosives is demonstrated at nanogram levels directly from surfaces in near real time including those that have complex geometries, those that are heat-sensitive, and those bearing complex sample matrices.

View Article and Find Full Text PDF

A benchtop miniature mass spectrometer system, Mini 12, with ambient ionization source and tandem mass spectrometry capabilities has been developed and characterized. This instrument was developed as a self-contained system to produce quantitative results for unprocessed samples of small volumes including nonvolatile analytes. The ion processing system, vacuum system, and control system are detailed.