A new method for the measurement of pulmonary gas exchange during inhalational anaesthesia is described which measures fresh gas and exhaust gas flows using carbon dioxide as an extractable marker gas. The theoretical precision of the method was compared by Monte Carlo modelling with other approaches which use marker gas dilution. A system was constructed for automated measurement of uptake of oxygen, nitrous oxide, volatile anaesthetic agent and elimination of carbon dioxide by an anaesthetized patient.
View Article and Find Full Text PDFObjective: The throughflow method is a new technique for continuous and minimally invasive measurement of cardiac output by the Fick principle, which uses ventilation of the 2 lungs with unequal inspired gas concentrations by means of a double-lumen endobronchial tube. It exploits steady-state gas exchange and thus permits rapid repetition of measurement.
Design: Comparison of paired measurements by the throughflow method using N(2)O exchange with bolus thermodilution.