Objective measurement of straylight in the human eye with a Shack–Hartmann (SH) wavefront aberrometer is limited in imaging angle. We propose a measurement principle and a point spread function (PSF) reconstruction algorithm to overcome this limitation. In our optical setup, a variable stop replaces the stop conventionally used to suppress reflections and scatter in SH aberrometers.
View Article and Find Full Text PDFGraefes Arch Clin Exp Ophthalmol
August 2013
Purpose: To examine the use of a modified Hartmann-Shack wave front aberrometer (WASCA; Carl Zeiss Meditec AG, Germany) to measure objective stray light caused by forward light scatter from the anterior segment of the human eye.
Setting: HELIOS Klinikum Erfurt/Institute of Biomedical Engineering and Informatics, Ilmenau University of Technology, Germany.
Method: Scatter parameters, including the Michelson contrast and cross-sectional area at half height (CAHM) were examined in Hartmann-Shack images from ten subjects with a cataract in one eye and an intraocular lens (IOL) in the other.