A reagent strip for quantifying theophylline in serum or plasma has been developed for use with the Apoenzyme Reactivation Immunoassay System (ARIS) with the Ames Seralyzer reflectance photometer. The test takes 80 s and involves comparison with a two-point calibration line, which can validly be stored in the instrument for two weeks. Results for theophylline in clinical serum samples correlate well (r greater than 0.
View Article and Find Full Text PDFThe Prostatic-Group-Label Immunoassay (PGLIA) technique has been incorporated into a reagent-strip format. We report use of flavin N6-(N'-2,4-dinitrophenyl-6-aminohexyl)adenine dinucleotide (DNP-FAD) as the prosthetic group derivative and 6-N-(2,4-dinitrophenyl)aminohexanoic acid (DNP-caproate) as the competing ligand. DNP-FAD not bound by antibody combines with glucose oxidase apoenzyme, which then reacts with glucose and oxygen, and gives color through a peroxidase-linked system.