Prediction of FRC using a respiratory P-V curve (2 1 syringe method) has been tested in eight patients with normal lungs and in 12 ARDS patients. FRC was measured using nitrogen dilution technique with a closed circuit. Correlation between measured and predicted FRC was excellent, especially when the expiratory limb of the P-V curve was used (r = 0.
View Article and Find Full Text PDFRoutine checking of the FIO2 of a mixture delivered by a ventilator revealed large discrepancies between the FIO2 assigned and that effectively delivered, the latter being low. Measurements of FIO2 at the wall outlets showed the O2 delivery pipeline to be contaminated by compressed air. By disconnecting all the ventilators supplied by this pipeline one after the other, one ventilator, or rather its blender, was found responsible.