Acta Anaesthesiol Scand
May 2015
Background: Case-by-case assessment of volatile anesthetic (VA) consumption is necessary to perform agent sparing dosing measures and for billing purposes. The gold standard technique for this purpose is to measure the weight difference of the vaporizer that occurs during VA delivery. Because suitable equipment is rarely available and weighing the vaporizer is only possible in prospectively planned fashion, a calculation method from recorded fresh gas flow (FGF) and VA courses is a viable alternative.
View Article and Find Full Text PDFJ Clin Monit Comput
October 2015
Knowledge of the consumed amount of volatile anesthetic (VA) expressed in liquid agent is necessary to enable agent sparing dosing measures and for billing purposes. The widespread Draeger Primus™ anesthesia machine displays in its logbook the amount of consumed VA at the end of each anesthesia, but the reliability of this parameter is yet unknown. The objective was to evaluate the precision and reliability of the inbuilt VA consumption display in Draeger Primus™ anesthesia machines as compared with the gold standard of weighing the vaporizer before and after anesthesia.
View Article and Find Full Text PDF