Publications by authors named "Nick Stroud"

Background: Little guidance exists regarding the minimum screw length and screw quantity necessary to achieve fixation in reverse total shoulder arthroplasty (rTSA); to that end, this study quantified the displacement of 2 different sizes of glenoid baseplates using multiple different screw lengths and quantities of screws in a low-density polyurethane bone-substitute model.

Methods: Testing of rTSA glenoid loosening was conducted according to ASTM F 2028-17. To independently evaluate the impact of screw quantity and screw length on rTSA glenoid fixation for 2 different sizes of glenoid baseplates, baseplates were constructed using 2 screws, 4 screws, or 6 screws (with the latter being used for the larger baseplate only) with 3 different poly-axial locking compression screw lengths.

View Article and Find Full Text PDF

The usage of and indications for total shoulder arthroplasty have grown in recent years. Certain aspects of these arthroplasty procedures can be very complex, especially in revision and fracture cases, often leading to proximal humerus bone loss. For cases with significant bone loss, there is a need for improved devices with additional options to treat a wider range of deformities while also mitigating existing complications and rates, such as poor distal fixation, inadequate soft tissue reattachment options, and joint instability.

View Article and Find Full Text PDF

Introduction: This study quantified glenoid fixation before and after cyclic loading of six reverse shoulder prosthesis designs when secured to low and high density bone substitute blocks.

Methods: A displacement test quantified fixation of six reverse shoulder designs: 38 mm Equinoxe standard offset (EQ), 38 mm Equinoxe lateral offset (EQL), 36 mm Depuy Delta III (DRS), 36 mm Zimmer (ZRS), 32 mm DJO RSP (DJO), and a 36 mm Tornier BIO-RSA (BIO), secured to 0.24 and 0.

View Article and Find Full Text PDF