Publications by authors named "Neil Crawford"

Background: Achieving optimal immediate stability is crucial in lumbar fusion surgeries. Traditionally, four pedicle screws have been utilized to provide posterior stability at the L5-S1 level. However, the use of bilateral transfacet pedicle screws (TFPS) as an alternative construct has shown promising results in terms of biomechanical stability.

View Article and Find Full Text PDF

Background: Synthetic computed tomography (sCT) can be created from magnetic resonance imaging (MRI) utilizing newer software. sCT is yet to be explored as a possible alternative to routine CT (rCT). In this study, rCT scans and MRI-derived sCT scans were obtained on a cadaver.

View Article and Find Full Text PDF
Article Synopsis
  • This study compared the biomechanics of cadaveric spines with mature fusions to normal spines in both intact and acutely fixed conditions.
  • Findings showed that mature fusion spines had significantly higher bone mineral density and less range of motion (ROM) compared to freshly fixed spines across various movements.
  • Additionally, adjacent levels in mature fusion spines displayed reduced mobility, particularly the levels immediately next to the fusion site, indicating potential impacts on overall spinal flexibility.
View Article and Find Full Text PDF

Study Design: Cadaveric study.

Objective: To evaluate accuracy, radiation exposure, and surgical time of a new robotic-assisted navigation (RAN) platform compared with freehand techniques in conventional open and percutaneous procedures.

Methods: Ten board-certified surgeons inserted 16 pedicle screws at T10-L5 (n = 40 per technique) in 10 human cadaveric torsos.

View Article and Find Full Text PDF

Study Design: Narrative review.

Objectives: Robotic systems in spinal surgery may offer potential benefits for both patients and surgeons. In this article, the authors explore the future prospects and current limitations of robotic systems in minimally invasive spine surgery.

View Article and Find Full Text PDF

Objective: Robotic spine surgery systems are increasingly used in the US market. As this technology gains traction, however, it is necessary to identify mechanisms that assess its effectiveness and allow for its continued improvement. One such mechanism is the development of a new 3D grading system that can serve as the foundation for error-based learning in robot systems.

View Article and Find Full Text PDF

Background: Biomechanical properties of intact spinal motion segments are used to establish baseline values during in vitro studies evaluating spinal surgical techniques and implants. These properties are also used to validate computational models (ie, patient-specific finite element models) of human lumbar spine segments. Our laboratory has performed a large number of in vitro mechanical studies of lumbar spinal segments, using a consistent methodology.

View Article and Find Full Text PDF

This study used a 3-dimensional (3D) craniocervical junction model of styloidogenic jugular venous compression (SJVC) syndrome to simulate and evaluate intracranial pressure (ICP) after internal jugular vein (IJV) compression by an elongated styloid process during axial rotation. The 3D-printed model created using data from an SJVC-syndrome patient included an articulating occipital-cervical junction, simplified arteriovenous system, gauge to measure simulated ICP, fixed obstruction simulating left-sided venous occlusion, and right-sided vascular tubing to simulate IJV compression. The model was rotated axially to its extreme right and left; maximum degree of motion and pressure were recorded for 3 cycles.

View Article and Find Full Text PDF

Background: In the emerging field of robot-assisted spine surgery, radiographic evaluation of pedicle screw accuracy in the surgical setting is of high interest. Advances in medical imaging have improved the accuracy of pedicle screw placement, from fluoroscopy-guided to computer-aided navigation.

Methods: A retrospective, institutional review board-exempt review of the first 106 navigated robot-assisted spine surgery cases was performed.

View Article and Find Full Text PDF
Article Synopsis
  • Robotic systems in surgery have advanced quickly, with da Vinci Surgical System installations in the U.S. nearly doubling from 2010 to 2017, especially for urological and gynecological procedures.
  • Adoption in spine surgery has been slower, but there is enthusiasm for robotic assistance due to the need for precision in complex, limited-access situations.
  • The FDA has approved 7 robotic systems for spinal surgery, and while clinical data indicates they are safe and accurate, further comparative studies are needed to fully understand their impact and improve broader adoption in this area.
View Article and Find Full Text PDF

Robotic assistance with integrated navigation is an area of high interest for improving the accuracy of minimally invasive pedicle screw placement. This study analyzes the accuracy of pedicle screw placement between an attending spine surgeon and a resident by comparing the left and right sides of the first 101 consecutive cases using navigated robotic assistance in a private practice clinical setting. A retrospective, Institutional Review Board-exempt review of the first 106 navigated robot-assisted spine surgery cases was performed.

View Article and Find Full Text PDF

Computer-aided navigation and robotic guidance systems have become widespread in their utilization for spine surgery. A recent innovation combines these two advances, which theoretically provides accuracy in spinal screw placement. This study describes the cortical and pedicle screw accuracy for the first 54 cases where navigated robotic assistance was used in a surgical setting.

View Article and Find Full Text PDF

Background: Information on the performance of posterior fixation with cortical screw (CS) versus pedicle screw (PS) trajectories for stabilizing thoracolumbar burst fractures is limited. Therefore, we sought to analyze stability with CS versus PS in short- and long-segment fixations using a 3-column spinal injury model.

Methods: Nondestructive flexibility tests: (1) intact, (2) intact + short fixation, (3) intact + long fixation, (4) after burst fracture, (5) short fixation + burst fracture, and (6) long fixation + burst fracture using thoracic spine segments (7 CS, 7 PS).

View Article and Find Full Text PDF

In the emerging field of robot-assisted spine surgery, the radiographic evaluation of pedicle screw accuracy in clinical application is an area of high interest. This study describes the pedicle screw accuracy of the first 56 consecutive cases in which navigated robotic assistance was used in a private practice clinical setting. A retrospective, Institutional Review Board-exempt review of the first 56 navigated robot-assisted spine surgery cases was performed.

View Article and Find Full Text PDF

There are potential pitfalls associated with the pursuit of accurate surgical navigation, such as vulnerability of the reference array to accidental dislodgment and damage or soiling of tracking arrays on tools. Additionally, there are hazards encountered when attempting accurate robotic screw placement in spine surgery, including skiving of the tool on bone, displacement of the robotic arm, or patient movement. Proven techniques are needed to address and mitigate these issues to ensure that navigation integrity is maintained and screw placement is accurate when using navigated robotic surgical guidance systems.

View Article and Find Full Text PDF

Objective: Lumbar interbody spacers are widely used in lumbar spinal fusion. The goal of this study is to analyze the biomechanics of a lumbar interbody spacer (Clydesdale Spinal System, Medtronic Sofamor Danek, Memphis, Tennessee, USA) inserted via oblique lumbar interbody fusion (OLIF) or direct lateral interbody fusion (DLIF) approaches, with and without posterior cortical screw and rod (CSR) or pedicle screw and rod (PSR) instrumentation.

Methods: Lumbar human cadaveric specimens (L2-L5) underwent nondestructive flexibility testing in intact and instrumented conditions at L3-L4, including OLIF or DLIF, with and without CSR or PSR.

View Article and Find Full Text PDF
Article Synopsis
  • Robotics in spinal surgery can enhance outcomes by reducing surgeon fatigue, improving screw accuracy, and minimizing radiation exposure.
  • The ExcelsiusGPS robotic system overcomes limitations of previous systems with features like a secure floor-mounted design, integrated neuronavigation, and real-time alerts for registration loss and skiving.
  • A case study highlights the successful application of ExcelsiusGPS in spinal surgery, demonstrating excellent screw placement and favorable patient outcomes with reduced radiation exposure.
View Article and Find Full Text PDF

Objective: We studied the effect of different cervicothoracic construct design variables on biomechanical stability in vitro.

Methods: Six fresh-frozen human cadaveric spines (C5-T4) were used. After intact analysis, each specimen was destabilized and reconstructed, with all groups having 4.

View Article and Find Full Text PDF

Objective: To evaluate the stability of multiple rod-connector construct designs using a mechanical 4-point bending testing frame.

Methods: A mechanical study was used to evaluate the bending stiffness of 3 connectors across 12 different configurations of rod-connector-rod constructs. Stability was evaluated in flexion-extension and lateral bending.

View Article and Find Full Text PDF

Background: The ExcelsiusGPS® (Globus Medical, Inc., Audubon, PA) is a next-generation spine surgery robotic system recently approved for use in the United States. The objective of the current study is to assess pedicle screw accuracy and clinical outcomes among two of the first operative cases utilizing the ExcelsiusGPS® robotic system and describe a novel metric to quantify screw deviation.

View Article and Find Full Text PDF

OBJECTIVEThere are limited data regarding the implications of revision posterior surgery in the setting of previous cervical arthroplasty (CA). The purpose of this study was to analyze segmental biomechanics in human cadaveric specimens with and without CA, in the context of graded posterior resection.METHODSFourteen human cadaveric cervical spines (C3-T1 or C2-7) were divided into arthroplasty (ProDisc-C, n = 7) and control (intact disc, n = 7) groups.

View Article and Find Full Text PDF

Background: Many approaches to the lumbar spine have been developed for interbody fusion. The biomechanical profile of each interbody fusion device is determined by the anatomical approach and the type of supplemental internal fixation. Lateral lumbar interbody fusion (LLIF) was developed as a minimally invasive technique for introducing hardware with higher profiles and wider widths, compared with that for the posterior lumbar interbody fusion (PLIF) approach.

View Article and Find Full Text PDF

The Excelsius GPS (Globus Medical, Inc.) was approved by the FDA in 2017. This novel robot allows for real-time intraoperative imaging, registration, and direct screw insertion through a rigid external arm-without the need for interspinous clamps or K-wires.

View Article and Find Full Text PDF

Objective: Recently developed expandable interbody spacers are widely accepted in spinal surgery; however, the resulting biomechanical effects of their use have not yet been fully studied. We analyzed the biomechanical effects of an expandable polyetheretherketone interbody spacer inserted through a bilateral posterior approach with and without different modalities of posterior augmentation.

Methods: Biomechanical nondestructive flexibility testing was performed in 7 human cadaveric lumbar (L2-L5) specimens followed by axial compressive loading.

View Article and Find Full Text PDF

Objective: To determine the stability of fusion constructs with unilateral pedicle screw (UPS) or bilateral pedicle screw (BPS) fixation with and without an interbody implant using the lateral lumbar interbody (LLIF) approach.

Methods: Standard nondestructive flexibility tests were performed on 13 cadaveric lumbar specimens to assess spinal stability of intact specimens and 5 configurations of posterior and interbody instrumentation. Spinal stability was determined as mean range of motion in flexion-extension, lateral bending, and axial rotation.

View Article and Find Full Text PDF