J Manipulative Physiol Ther
December 2024
Objective: The purpose of this study was to examine the intrarater reliability and criterion validity of a digital goniometer for measuring cervical range of motion including an active sitting version of the cervical flexion rotation test, an important clinical measure for capturing upper cervical dysfunction.
Methods: A cross-sectional study (n = 18; 78% women) examined total cervical and upper cervical active ranges of motion (flexion, extension, side bending, rotation) measured concurrently using 3 measurement methods (a digital goniometer, the cervical range-of-motion instrument, and 3-dimensional motion analysis). Intraclass correlation coefficient (2,1), Pearson correlation coefficients (digital goniometer versus 3-dimensional motion analysis), and minimum detectable change were calculated.