Publications by authors named "N P Ferris"

Article Synopsis
  • Magnetic fields that switch at high frequencies (kHz) can cause peripheral nerve stimulation (PNS) in the body, and this has primarily been studied below 10 kHz.
  • Previous studies above ~25 kHz showed inconsistent results with high variability, but new research involving 8 volunteers measured PNS at frequencies from 200 Hz to 88.1 kHz.
  • Results indicated that instead of continuously decreasing, the PNS thresholds hit a minimum around 25 kHz and then increased by 39% from that point to 88.1 kHz, highlighting the need for more research on the biological mechanisms behind magnetostimulation at these higher frequencies.
View Article and Find Full Text PDF

Advances in the spatiotemporal resolution and field-of-view of neuroimaging tools are driving mesoscale studies for translational neuroscience. On October 10, 2023, the Center for Mesoscale Mapping (CMM) at the Massachusetts General Hospital (MGH) Athinoula A. Martinos Center for Biomedical Imaging and the Massachusetts Institute of Technology (MIT) Health Sciences Technology based Neuroimaging Training Program (NTP) hosted a symposium exploring the state-of-the-art in this rapidly growing area of research.

View Article and Find Full Text PDF

Rapid switching of magnetic resonance imaging (MRI) gradient fields induces electric fields that can cause peripheral nerve stimulation (PNS) and so accurate characterization of PNS is required to maintain patient safety and comfort while maximizing MRI performance. The minimum magnetic gradient amplitude that causes stimulation, the PNS threshold, depends on intrinsic axon properties and the spatial and temporal properties of the induced electric field. The PNS strength-duration curve is widely used to characterize simulation thresholds for periodic waveforms and is parameterized by the chronaxie and rheobase.

View Article and Find Full Text PDF

Purpose: Peripheral nerve stimulation (PNS) limits the usability of state-of-the-art whole-body and head-only MRI gradient coils. We used detailed electromagnetic and neurodynamic modeling to set an explicit PNS constraint during the design of a whole-body gradient coil and constructed it to compare the predicted and experimentally measured PNS thresholds to those of a matched design without PNS constraints.

Methods: We designed, constructed, and tested two actively shielded whole-body Y-axis gradient coil winding patterns: YG1 is a conventional symmetric design without PNS-optimization, whereas YG2's design used an additional constraint on the allowable PNS threshold in the head-imaging landmark, yielding an asymmetric winding pattern.

View Article and Find Full Text PDF

Purpose: Modern high-amplitude gradient systems can be limited by the International Electrotechnical Commission 60601-2-33 cardiac stimulation (CS) limit, which was set in a conservative manner based on electrode experiments and E-field simulations in uniform ellipsoidal body models. Here, we show that coupled electromagnetic-electrophysiological modeling in detailed body and heart models can predict CS thresholds, suggesting that such modeling might lead to more detailed threshold estimates in humans. Specifically, we compare measured and predicted CS thresholds in eight pigs.

View Article and Find Full Text PDF