Cardiovasc Intervent Radiol
December 2014
Purpose: Devices constantly tracking the blood pressure (BP) of hypertensive patients are highly desired to facilitate effective patient management and to reduce hospitalization. We report on experiences gathered in a pilot study that was designed to evaluate the prototype of a newly developed, minimally invasive implantable sensor system for long-term BP monitoring.
Methods: The device was implanted in the femoral artery (FA) of 12 sheep via standard FA catheterization under fluoroscopic control.
IEEE Trans Biomed Eng
November 2012
This paper presents a novel fully implantable wireless sensor system intended for long-term monitoring of hypertension patients, designed for implantation into the femoral artery with computed tomography angiography. It consists of a pressure sensor and a telemetric unit, which is wirelessly connected to an extracorporeal readout station for energy supply and data recording. The system measures intraarterial pressure at a sampling rate of 30 Hz and an accuracy of ±1.
View Article and Find Full Text PDF