We describe a gold nanoparticle-based sandwich immunoassay for the dual detection and measurement of hemoglobin A1c (HbA1c) and total hemoglobin in the whole blood (without pretreatment) in a single step for personalized medicine. The optimized antibody-functionalized gold nanoparticles immunoreact simultaneously with HbA1c and total hemoglobin to form a sandwich at distinctive test lines to transduce visible signals. The applicability of this method as a personal management tool was demonstrated by establishing a calibration curve to relate % HbA1c, a useful value for type 2 diabetes management, to the signal ratio of captured HbA1c to all other forms of hemoglobin.