The present study is aimed at exploring different scanning parameters, detectors and their orientations for time-efficient and accurate commissioning of a 6 MV clinical linear accelerator (LINAC). Beam profiles and percentage depth dose (PDD) curves were measured with a PTW dosimetry diode, a PTW Semiflex and a PinPoint ion chamber in different orientations. To acquire beam data, equidistant (step size of 0.
View Article and Find Full Text PDF