Publications by authors named "Morteza Ebrahimzadeh Torkamani"

Purpose: Whole-body photoneutron (PN) energy-specific dosimetry on phantom surfaces and 360°angular fast PN distribution dosimetry on and around polyethylene (PE) phantom organs for prostate cancer in 10 cm × 10 cm and 20 cm × 20 cm field sizes of 18 MV X-rays of Varian Clinac 2100C linear accelerator.

Methods: Novel "Sohrabi neutron dosimetry methods" including "miniature passive neutron dosimeter/spectrometer" and "strip polycarbonate neutron dosimeters" were applied. Energy-specific PN dose equivalents on surface as well as 360°fast PN dose equivalent angular distributions on and around PE phantom organs; head, neck, thorax, arms, pelvis, thighs and legs were determined.

View Article and Find Full Text PDF

Breakthrough whole body energy-specific photoneutron (PN) dosimetry was made in/out-of-field in polyethylene phantom organ surface/depths remote from isocenter of 10 × 10 cm field prostate cancer therapy in 18 MV X-rays Varian Clinac 2100C medical linear accelerator for PN tissue-specific second primary cancer (PN-SPC) risk estimation. A novel miniature neutron dosimeter/spectrometer with polycarbonate/B/cadmium inserts was invented and applied. Each dosimeter determines seven tissue-specific dose equivalent (mSv)/Gy X-ray dose at each measurement point providing seven major energy-specific responses for beam thermal, albedo thermal, total thermal, total epithermal, total fast, sum of totals (thermal + epithermal) and sum of totals (thermal + epithermal + fast) PNs dose equivalents.

View Article and Find Full Text PDF