We report the case of a 72-year-old patient with ischemic cardiomyopathy who was found to have a lead fracture in the RV coil DF-1 connector of an integrated bipolar implantable cardioverter-defibrillator (ICD) lead system at the time of upgrade to a biventricular ICD. Due to the redundant connection of the RV coil to the DF-1 connector and the ring on the IS-1 connector, a custom-designed adapter was used to exclude the site of fracture. This novel solution spared the patient the potential risks of lead extraction or abandoning the fractured ICD lead.

View Article and Find Full Text PDF

Objective: To evaluate the safety and efficacy of using a circular multielectrode catheter for mapping and ablation of ventricular tachycardia (VT) or premature ventricular complexes (PVCs) from the right ventricular outflow tract (RVOT).

Background: Three-dimensional (3D) mapping systems are commonly used for mapping and ablation of RVOT VT and PVCs. Newer catheters that are circular with multiple electrodes, such as the Lasso catheter, are capable of simultaneously recording from multiple points within a circumferential plane.