Publications by authors named "Minyi Ye"

Blood lactate and blood pressure measurements are important predictors of life-threatening complications after infant open-heart surgeries requiring cardiopulmonary bypass (CPB). We have developed an intravascular nitric oxide (NO)-releasing 5-Fr catheter that contains a lactate sensor for continuous in-blood lactate monitoring and a dedicated lumen for third-party pressure sensor attachment. This device has antimicrobial and antithrombotic properties and can be implanted intravascularly.

View Article and Find Full Text PDF

We present an animal model used to evaluate the in vivo performance of electrochemical amperometric continuous lactate sensors compared to blood gas instruments. Electrochemical lactate sensors were fabricated, placed into 5 Fr central venous catheters (CVCs), and paired with wireless potentiostat devices. Following in vivo evaluation and calibration, sensors were placed within the jugular and femoral veins of a porcine subject as a preliminary assessment of in vivo measurement accuracy.