Purpose: This report offers a detailed examination of the inception and current state of the Heavy-ion Therapy Center (HITC) at the Yonsei Cancer Center (YCC), setting it apart as the world's first center equipped with a fixed beam and two superconducting gantries for carbon-ion radiation therapy (CIRT).
Materials And Methods: Preparations for CIRT at YCC began in 2013; accordingly, this center has completed a decade of meticulous planning and culminating since the operational commencement of the HITC in April 2023.
Results: This report elaborates on the clinical preparation for adopting CIRT in Korea.
Purpose: This study presents novel quality assurance (QA) approach for volumetric modulated arc therapy (VMAT) that leverages frame-by-frame electronic portal imaging device (EPID) images integrated into Mobius3D for accurate three-dimensional dose calculations.
Methods: Sequential EPID images for VMAT plans were acquired every 0.4-second by iView system and processed through iterative deconvolution to mitigate blurring from photon scattering.
In breast cancer radiation therapy, minimizing radiation-related risks and toxicity is vital for improving life expectancy. Tailoring radiotherapy techniques and treatment positions can reduce radiation doses to normal organs and mitigate treatment-related toxicity. This study entailed a dosimetric comparison of six different external beam whole-breast irradiation techniques in both supine and prone positions.
View Article and Find Full Text PDFPurpose: This study reviewed and meta-analyzed evidence on radiomics-based hybrid models for predicting radiation pneumonitis (RP). These models are crucial for improving thoracic radiotherapy plans and mitigating RP, a common complication of thoracic radiotherapy. We examined and compared the RP prediction models developed in these studies with the radiomics features employed in RP models.
View Article and Find Full Text PDFThis work aims to investigate the clinical feasibility of deep learning-based synthetic CT images for cervix cancer, comparing them to MR for calculating attenuation (MRCAT). Patient cohort with 50 pairs of T2-weighted MR and CT images from cervical cancer patients was split into 40 for training and 10 for testing phases. We conducted deformable image registration and Nyul intensity normalization for MR images to maximize the similarity between MR and CT images as a preprocessing step.
View Article and Find Full Text PDFWe aimed to investigate the deliverability of dynamic conformal arc therapy (DCAT) by gantry wobble owing to the intrinsic inter-segment break of the Elekta linear accelerator (LINAC) and its adverse influence on the dose to the patient. The deliverability of DCAT was evaluated according to the plan parameters, which affect the gantry rotation speed and resultant positional inaccuracies; the deliverability according to the number of control points and dose rates was investigated by using treatment machine log files and dosimetry devices, respectively. A non-negligible degradation in DCAT deliverability due to gantry wobble was observed in both the treatment machine log files and dosimetry devices.
View Article and Find Full Text PDFPurpose: The aim of this work was to prospectively evaluate the toxicity and cosmetic outcomes of 5-fraction, stereotactic, accelerated partial breast irradiation (APBI).
Methods And Materials: This prospective observational cohort study enrolled women who underwent APBI for invasive carcinoma or carcinoma in situ of the breast. APBI was delivered using a CyberKnife M6 robotic radiosurgery system at 30 Gy in 5 nonconsecutive, once-daily fractions.
. To address the shortcomings of current procedures for evaluating the measured-to-planned dose agreement indosimetry (IVD), this study aimed to develop an accurate and efficient novel framework to identify the detector location placed on a patient's skin surface using a 3D camera and determine the planned dose at the same anatomical position corresponding to the detector location..
View Article and Find Full Text PDFTechnol Cancer Res Treat
December 2022
We aimed to investigate the dose calculation accuracy of Mobius3D for small-field flattening-filter-free x-rays, mainly utilized for stereotactic body radiation therapy (SBRT). The accuracy of beam modeling and multileaf collimator (MLC) modeling in Mobius3D, significantly affecting the dose calculation is investigated. The commissioning procedures of Mobius3D were performed for unflattened 6 MV and 10 MV x-ray beams of the linear accelerator, including beam model adjustment and dosimetric leaf gap (DLG) optimization.
View Article and Find Full Text PDFThis work attempted to construct a new metal artifact reduction (MAR) framework in kilo-voltage (kV) computed tomography (CT) images by combining (1) deep learning and (2) multi-modal imaging, defined as MARTIAN (Metal Artifact Reduction throughout Two-step sequentIAl deep convolutional neural Networks). Most CNNs under supervised learning require artifact-free images to artifact-contaminated images for artifact correction. Mega-voltage (MV) CT is insensitive to metal artifacts, unlike kV CT due to different physical characteristics, which can facilitate the generation of artifact-free synthetic kV CT images throughout the first network (Network 1).
View Article and Find Full Text PDFBackground: Based on a previous study on the feasibility of TomoEQA, an exit detector-based patient-specific pre-treatment quality assurance (QA) method for helical tomotherapy, an in-depth clinical evaluation was conducted.
Methods: Data of one hundred patients were analyzed to evaluate the clinical usefulness of TomoEQA for patient-specific pre-treatment QA in comparison with the conventional phantom-based method. Additional investigations were also performed under unusual measurement conditions to validate the off-axis region.
Various deformable image registration (DIR) methods have been used to evaluate organ deformations in 4-dimensional computed tomography (4D CT) images scanned during the respiratory motions of a patient. This study assesses the performance of 10 DIR algorithms using 4D CT images of 5 patients with fiducial markers (FMs) implanted during the postoperative radiosurgery of multiple lung metastases. To evaluate DIR algorithms, 4D CT images of 5 patients were used, and ground-truths of FMs and tumors were generated by physicians based on their medical expertise.
View Article and Find Full Text PDFJ Appl Clin Med Phys
February 2022
Purpose: The purpose of this study was to investigate the matching error that occurs when the Mobius3D fingerprinting system is applied in conjunction with an Elekta linear accelerator (LINAC) and to offer an acceptable and alternative method for circumventing this problem.
Material And Methods: To avoid the multileaf collimator (MLC) conflicting error in the Mobius3D fingerprinting system, we developed an in-house program to move the MLC in the Digital Imaging and Communications in Medicine (DICOM) radiotherapy (RT)-Plan to pertinent positions, considering the relationship between log data and planned data. The re-delivered log files were calculated in the Mobius3D system, and the results were compared with those of corrected data (i.
Purpose: To develop an internal target volume (ITV) margin determination framework (or decision-supporting framework) for treating multiple lung metastases using CyberKnife Synchrony with intraoperatively implanted fiducial markers (IIFMs). The feasibility of using non-ideally implanted fiducial markers (a limited number and/or far from a target) for tracking-based lung stereotactic ablative radiotherapy (SABR) was investigated.
Methods: In the developed margin determination framework, an optimal set of IIFMs was determined to minimize a tracking uncertainty-specific ITV (ITV) margin (margin required to cover target-to-marker motion discrepancy), i.
Purpose: Tetrahedral mesh (TM)-based computational human phantoms have recently been developed for evaluation of exposure dose with the merit of precisely representing human anatomy and the changing posture freely. However, conversion of recently developed TM phantoms to the Digital Imaging and Communications in Medicine (DICOM) file format, which can be utilized in the clinic, has not been attempted. The aim of this study was to develop a technique, called TET2DICOM, to convert the TM phantoms to DICOM datasets for accurate treatment planning.
View Article and Find Full Text PDFParticle therapy is a promising and evolving modality of radiotherapy that can be used to treat tumors that are radioresistant to conventional photon beam radiotherapy. It has unique biological and physical advantages compared with conventional radiotherapy. The characteristic feature of particle therapy is the "Bragg peak," a steep and localized peak of dose, that enables precise delivery of the radiation dose to the tumor while effectively sparing normal organs.
View Article and Find Full Text PDFPurpose: How modern cardiac sparing techniques and beam delivery systems using advanced x-ray and proton beam therapy (PBT) can reduce incidental radiation exposure doses to cardiac and pulmonary organs individually or in any combination is poorly investigated.
Methods: Among 15 patients with left-sided breast cancer, partial wide tangential 3D-conformal radiotherapy (3DCRT) delivered in conventional fractionation (CF) or hypofractionated (HF) schedules; PBT delivered in a CF schedule; and volumetric modulated arc therapy (VMAT) delivered in an HF schedule, each under continuous positive airway pressure (CPAP) and free-breathing (FB) conditions, were examined. Target volume coverage and doses to organs-at-risk (OARs) were calculated for each technique.
This study evaluates the changes occurring in the X-ray energy of a linear accelerator (LINAC) using a Daily QA3 detector system. This is accomplished by comparing the Daily QA3 results against those obtained using a water phantom. The X-energy levels of a LINAC were monitored over a duration of 1 month using the Daily QA3 system.
View Article and Find Full Text PDFPurpose: Existing phantom-less quality assurance (QA) platforms does not provide patient-specific QA for helical tomotherapy (HT). A new system, called TomoEQA, is presented to facilitate this using the leaf open time (LOT) of a binary multi-leaf collimator, as measured by an exit detector.
Methods: TomoEQA was designed to provide measurement-based LOTs based on detector data and to generate a new digital imaging and communication in medicine (DICOM) dataset that includes the measured LOTs for use by secondary check platforms.
Purpose: This study aimed to investigate the parameters with a significant impact on delivery quality assurance (DQA) failure and analyze the planning parameters as possible predictors of DQA failure for helical tomotherapy.
Methods: In total, 212 patients who passed or failed DQA measurements were retrospectively included in this study. Brain (n = 43), head and neck (n = 37), spinal (n = 12), prostate (n = 36), rectal (n = 36), pelvis (n = 13), cranial spinal irradiation and a treatment field including lymph nodes (n = 24), and other types of cancer (n = 11) were selected.
Background: This study investigated the effect of accumulated doses on radio-photoluminescence glass dosimeters (RPLGDs) from measurements involving mega-voltage photons.
Methods: Forty-five commercially available RPLGDs were irradiated to estimate their dose responses. Photon beams of 6, 10, and 15 MV were irradiated onto the RPLGDs inside a phantom, which were divided into five groups with different doses and energies.
The European Society of Radiation & Oncology and Advisory Committee on Radiation Oncology Practice (ESTRO-ACROP) presented new guidelines for clinical target volume (CTV) delineation in post-mastectomy radiation therapy (PMRT) after implant-based immediate breast reconstruction (IBR-i). This study evaluated the dosimetric characteristics, dosimetric accuracy, and delivery accuracy of these guidelines in volumetric modulated arc therapy (VMAT). This retrospective study included 15 patients with left breast cancer who underwent mastectomy with tissue expander placement followed by PMRT.
View Article and Find Full Text PDFPurpose: To provide practical guidelines for Mobius3D commissioning based on experiences of commissioning/clinical implementation of Mobius3D and MobiusFX as patient-specific quality assurance tools on multiple linear accelerators.
Methods: The vendor-suggested Mobius3D commissioning procedures, including beam model adjustment and dosimetric leaf gap (DLG) optimization, were performed for 6 MV X-ray beams of six Elekta linear accelerators. For the beam model adjustment, beam data, such as the percentage depth dose, off-axis ratio (OAR), and output factor (OF), were measured using a water phantom and compared to the vendor-provided reference values.