Introduction: Clinical evaluation of a two-dimensional (2D) liquid-filled ion chamber detector array used in the verification of highly modulated small beams of stereotactic body radiation therapy (SBRT) has been conducted.
Materials And Methods: Measurements with the Octavius 1000 SRS (PTW, Freiburg, Germany) detector with 977 liquid-filled ion chambers were compared against EDR2 film and PTW Octavius Seven29. The performance of detector array has been evaluated on ten SBRT patient plans.
Methods: Five patients with 38 fields have been analyzed in this study. The plans were optimized for the following clinical sites: one liver, one lung, one brain, one prostate and one spine. The detector array used for the measurements was the PTW Seven29 array.
View Article and Find Full Text PDFPurpose: The purpose of the study is to investigate the characteristics of a two-dimensional (2D) liquid-filled ion chamber detector array, which is used for the verification of radiotherapy treatment plans that use small field sizes of up to 10 × 10 cm.
Methods: The device used in this study was Octavius 1000 SRS model (PTW, Freiburg, Germany). Its 2D array of detectors consists of 977 liquid-filled ion chambers arranged over an area of 11 × 11 cm.