Purpose: To determine if noncontrast T1-weighted (T1W) images from 3T magnetic resonance (MR) imaging accurately depict radiofrequency (RF) ablation zones as determined macroscopically and microscopically in a blood-perfused bovine liver model.
Materials And Methods: Three-dimensional (3D) gradient-recalled echo (GRE) T1W images were obtained on a 3T MR imaging scanner after RF ablations (n = 14) of in vitro blood-perfused bovine livers. The resulting central hypointense and peripheral hyperintense signal regions were measured and compared with the inner tan and outer red zones of the gross specimen.
Purpose: To describe an in vitro blood-perfused bovine liver model for the testing of radiofrequency (RF) ablation devices and compare the performance of a specific RF ablation device in the model relative to three other biologic models.
Materials And Methods: Fresh bovine livers were used to create three in vitro models: blood-perfused, Krebs-Henseleit (KH) solution-perfused, and nonperfused. The perfused models were connected to a heart-lung machine via the portal vein and perfused with heparinized autologous blood or KH solution under physiologic conditions.