We conducted a radiological safety and quality assurance (QA) audit of 118 medical X-ray diagnostic machines installed in 45 major hospitals in India. The main objective of the audit was to verify compliance with the regulatory requirements stipulated by the national regulatory body. The audit mainly covered accuracy check of accelerating potential (kVp), linearity of tube current (mA station) and timer, congruence of radiation and optical field, and total filtration; in addition, we also reviewed medical X-ray diagnostic installations with reference to room layout of X-ray machines and conduct of radiological protection survey.

View Article and Find Full Text PDF

Exposure of children to ionising radiation is considered to carry higher risk than that of adults; therefore a need to suggest diagnostic reference levels (DRLs) for the common paediatric diagnostic X-ray procedures was recognised for the X-ray machines meeting the requirements of the recently implemented Safety Code for Medical Diagnostic X-ray Equipment and Installations in India. Measurements were carried out for entrance surface air kerma (free in air) in conventional paediatric X-ray diagnostic examinations among four age groups: <1, 1-4, 5-9 and 10-15 y. A total of 2240 air kerma measurements at different fixed focus to skin distances were studied for 7 paediatric diagnostic examinations with 11 different projections on 62 X-ray machines installed in 22 selected hospitals in the country.