In this work, overshoot and undershoot effects associated with step-and-shoot IMRT (SSIMRT) delivery on a Varian Clinac 21iX are investigated, and their impact on patient-specific QA point dose measurements and treatment plan delivery are evaluated. Pinnacle3 SSIMRT plans consisting of 5, 10, and 15 identical 5 × 5 cm2 MLC defined segments and MU/segment values of 5 MU, 10 MU, and 20 MU were utilized and delivered at 600/300 MU/min. Independent of the number of segments the overshoot and undershoot at 600 MU/min were approximately ± 10%, ± 5%, and ± 2.