The U.S. Navy uses the Harshaw 8840/8841 dosimetric (DT-702/PD) system, which employs LiF:Mg,Cu,P thermoluminescent dosimeters (TLDs), developed and produced by Thermo Fisher Scientific (TFS).
View Article and Find Full Text PDFThe United States Navy monitors the dose its radiation workers receive using the DT-702/PD thermoluminescent dosimeter, which consists of the Harshaw 8840 holder and the four-element Harshaw 8841 card. There were two main objectives of this research. In the first objective, the dosimeters were exposed to 100 Gy using electron and x-ray beams and found to respond approximately 30-40% lower than the delivered dose.
View Article and Find Full Text PDFA short-term sensitivity loss in LiF:Mg,Cu,P thermoluminescent dosemeters (TLDs) was observed and is described. Its observation occurred during a pre-irradiation anneal with a slightly elevated maximum temperature (5-15°C), which causes notable under-response (5-10 %) of the subsequent read at the recommended time-temperature profile (TTP), which has a peak temperature of 260°C. A subsequent irradiation and reading using the recommended TTP showed partial or complete recovery of the TLD's sensitivity.