Pharmaceuticals reportedly cause damage to some polymeric medical devices that administer them. Because this phenomenon and its causes still remain unclear, in this study, all the possible combinations of polymeric materials and pharmaceutical ingredients that could cause failures were identified by conducting a comprehensive analysis on a wide variety of such combinations and through verification tests using the products. The results of the simple immersion tests and the reports of clinical failures indicated that the failures were not caused by the lack of chemical resistance of the polymers but by the environmental stress cracking (ESC) induced by a combination of the stress generated in the material and the interaction with a specific chemical.
View Article and Find Full Text PDFAmbulatory blood pressure monitoring (ABPM) is used for the evaluation of out-of-office blood pressure (BP), however, knowledge concerning the detailed behavior of nocturnal blood pressure (BP) and pulse rate (PR) is limited.A total of 190 participants (64 ± 15 years, 46.3% males) underwent ABPM for diagnosis of hypertension or evaluation of hypertensive therapy.