Publications by authors named "Marijke Ikink"

We characterized the on-board megavoltage imager (MVI) of a magnetic resonance-guided radiotherapy machine for beam output checks. Linearity and repeatability of its dose response were investigated. Alignment relative to the beam under clinical circumstances was evaluated for a year using daily measurements.

View Article and Find Full Text PDF

Beam output checks of a commercial high-field magnetic resonance-guided radiotherapy machine can be performed with its on-board megavoltage imager (MVI). This is a fast and efficient method, but only recommended for daily checks. The aim of our study was to show its suitability for weekly checks by investigating its long-term agreement with the golden standard: ionization chamber measurements in a water tank.

View Article and Find Full Text PDF

The purpose of this study was to develop and evaluate filmless quality assurance (QA) tools for helical tomotherapy by using the signals from the on-board megavoltage computed tomography (MVCT) detector and applying a dedicated step-wedge phantom. The step-wedge phantom is a 15 cm long step-like aluminum block positioned on the couch. The phantom was moved through the slit beam and MVCT detector signals were analyzed.

View Article and Find Full Text PDF