Objectives: Monoclonal immunoglobulins can cause interference in many laboratory analyses. During a 4 month period we observed seven patients with monoclonal disease and falsely extremely elevated 25-hydroxyvitamin D (25(OH)D) results above 160 ng/mL (>400 nmol/L) measured using an immunoassay from Abbott Diagnostics. Based on these findings, we studied the occurrence of falsely elevated 25(OH)D in samples with paraproteins and investigated possible mechanisms of the observed interference.
View Article and Find Full Text PDF