The present study aimed to validate the accuracy of blood pressure (BP) measurements of the OMRON M3500 automatic oscillometric BP device by using normal- and high-speed modes in adult and specific populations (obese and children) according to American National Standards Institute/Association for the Advancement of Medical Instrumentation (AAMI)/International Organization for Standardization/81060-2:2009 protocol. The validation was performed in 135 patients: 100 adults and 35 children (aged 3-12 years). The results showed a mean error of the individual paired determinations of the OMRON device and mercury sphygmomanometers within values of criterion 1 (5±8 mm Hg).