Purpose: The proximity or overlap of planning target volume (PTV) and organs-at-risk (OARs) poses a major challenge in stereotactic body radiation therapy (SBRT) of pancreatic cancer (PACA). This international treatment planning benchmark study investigates whether simultaneous integrated boost (SIB) and simultaneous integrated protection (SIP) concepts in PACA SBRT can lead to improved and harmonized plan quality.
Methods And Materials: A multiparametric specification of desired target doses (gross target volume [GTV], GTV, PTV, and PTV) with 2 prescription doses of GTV = 5 × 9.
This topical review focuses on Patient-Specific Quality Assurance (PSQA) approaches to stereotactic body radiation therapy (SBRT). SBRT requires stricter accuracy than standard radiation therapy due to the high dose per fraction and the limited number of fractions. The review considered various PSQA methods reported in 36 articles between 01/2010 and 07/2022 for SBRT treatment.
View Article and Find Full Text PDFPurpose: To determine the error detection sensitivity of a commercial log file-based system (LINACWatch®, LW) for integration into clinical routine and to compare it with a measurement device (OCTAVIUS 4D, Oct4D) for IMRT and VMAT delivery QA.
Materials And Methods: 76 VMAT/IMRT plans (H&N, prostate, rectum and breast) preliminarily classified according to their Modulation Complexity Score (MCS) calculated by LW, were considered. Receiver Operating Characteristic (ROC) Curves were used to establish gamma criteria for LW.
Purpose: The influence of basic plan parameters such as slice thickness, grid resolution, algorithm type and field size on calculated small field output factors (OFs) was evaluated in a multicentric study.
Methods And Materials: Three computational homogeneous water phantoms with slice thicknesses (ST) 1, 2 and 3 mm were shared among twenty-one centers to calculate OFs for 1x1, 2x2 and 3x3 cm field sizes (FSs) (normalized to 10x10 cm FS), with their own treatment planning system (TPS) and the energy clinically used for stereotactic body radiation therapy delivery. OFs were calculated for each combination of grid resolution (GR) (1, 2 and 3 mm) and ST and finally compared with the OFs measured for the TPS commissioning.
Objectives: This multicentric study was carried out to investigate the impact of small field output factors (OFs) inaccuracies on the calculated dose in volumetric arctherapy (VMAT) radiosurgery brain plans.
Methods: Nine centres, realised the same five VMAT plans with common planning rules and their specific clinical equipment Linac/treatment planning system commissioned with their OFs measured values (OFbaseline). In order to simulate OFs errors, two new OFs sets were generated for each centre by changing only the OFs values of the smallest field sizes (from 3.
Background: COVID-19 in Italy has led to the need to reorganize hospital protocols with a significant risk of interruption to cancer treatment programs. In this report, we will focus on a management model covering the two phases of the COVID-19 emergency, namely lockdown-phase I and post-lockdown-phase II.
Methods: The following steps were taken in the two phases: workload during visits and radiotherapy planning, use of dedicated routes, measures for triage areas, management of suspected and positive COVID-19 cases, personal protective equipment, hospital environments and intra-institutional meetings and tumor board management.
Background And Purpose: High-dose 6‑MV radiotherapy may cause cardiac implantable electronic devices (CIEDs) to malfunction. To assess CIED malfunctions resulting from direct exposure up to 10 Gy, 100 pacemakers (PMs) and 40 implantable cardioverter-defibrillators (ICDs) were evaluated.
Materials And Methods: CIEDs underwent baseline interrogation.
Stereotactic body radiation therapy (SBRT) has been recognized as a standard treatment option for many anatomical sites. Sophisticated radiation therapy techniques have been developed for carrying out these treatments and new quality assurance (QA) programs are therefore required to guarantee high geometrical and dosimetric accuracy. This paper focuses on recent advances on in-vivo measurements methods (IVM) for SBRT treatment.
View Article and Find Full Text PDFPurpose: Left ventricular assist device (LVAD) is considered a standard care for patients with advanced heart failure. The aim of this work was to study in vitro the effects of direct exposure of the Jarvik 2000 LVAD to 10-MV photon beams.
Methods: Jarvik 2000 pump was immersed in a siliconized box filled with deionized water.
Purpose: The complexity of the modern Stereotactic Body Radiation Therapy (SBRT) techniques requires comprehensive quality assurance programs, to ensure the right treatment to the patient. Dosimetry of small radiation fields is a challenge especially for radiotherapy centres starting to work on this issue. The matter to be discussed here concerns the need of detailed measurement procedures and cross checks to be paired to the usual recommendations on detectors and correction factors.
View Article and Find Full Text PDFTechnol Cancer Res Treat
January 2018
Purpose: The aim of the in vivo dosimetry, during the fractionated radiation therapy, is the verification of the correct dose delivery to patient. Nowadays, in vivo dosimetry procedures for photon beams are based on the use of the electronic portal imaging device and dedicated software to elaborate electronic portal imaging device images.
Methods: In total, 8474 in vivo dosimetry tests were carried out for 386 patients treated with 3-dimensional conformal radiotherapy, intensity-modulated radiotherapy, and volumetric modulated arc therapy techniques, using the SOFTDISO.
Interventional radiology is based on minimally invasive procedures that allow diagnosis and percutaneous treatment of diseases in almost all organ systems. Such procedures have many benefits, but they also contribute significantly to collective radiation dose. In this regard, effective dose (E) is a convenient quantity to estimate patients' stochastic radiation risk.
View Article and Find Full Text PDFPurpose: The aim of the study was a multicenter evaluation of MLC&jaws-defined small field output factors (OF) for different linear accelerator manufacturers and for different beam energies using the latest synthetic single crystal diamond detector commercially available. The feasibility of providing an experimental OF data set, useful for on-site measurements validation, was also evaluated.
Methods: This work was performed in the framework of the Italian Association of Medical Physics (AIFM) SBRT working group.
This study aims at optimizing treatment planning in young patients affected by lymphoma (Stage II to III) by using an inclined board (IB) that allows reducing doses to the organs at risk. We evaluated 19 young patients affected by stage I to III lymphomas, referred to our Department for consolidation radiotherapy (RT) treatment on the mediastinum. Patients underwent 2 planning computed tomography (CT) scans performed in different positions: flat standard position and inclined position.
View Article and Find Full Text PDFPurpose: In prostate cancer radiotherapy, the relationship between genitourinary (GU) toxicity and clinical-dosimetric parameters is debated. We report our analysis of the parameters associated with GU toxicity.
Materials And Methods: Eighty-six consecutive patients treated with conformal radiotherapy for localized prostate cancer were retrospectively analyzed; the bladder was delineated both as "whole bladder" (WB: Defined in its entirety as a solid organ) and "inferior bladder" (IB: Corresponding to the distal part of the bladder).
Purpose: A large-scale multi-institutional planning comparison on lung cancer SABR is presented with the aim of investigating possible criticism in carrying out retrospective multicentre data analysis from a dosimetric perspective.
Methods: Five CT series were sent to the participants. The dose prescription to PTV was 54Gy in 3 fractions of 18Gy.
Purpose: The aim of the present work was to evaluate small field size output factors (OFs) using the latest diamond detector commercially available, PTW-60019 microDiamond, over different CyberKnife systems. OFs were measured also by silicon detectors routinely used by each center, considered as reference.
Methods: Five Italian CyberKnife centers performed OFs measurements for field sizes ranging from 5 to 60mm, defined by fixed circular collimators (5 centers) and by Iris(™) variable aperture collimator (4 centers).
The purpose of this study was to compare the dosimetric properties of small field electron beams shaped by circular Cerrobend blocks and stainless steel tubular applicators. Percentage depth dose curves, beam profiles, and output factors of small-size circular fields from 2 to 5 cm diameter, obtained either by tubular applicators and Cerrobend blocks, were measured for 6, 10, and 15 MeV electron beam energies. All measurements were performed using a PTW microDiamond 60019 premarket prototype.
View Article and Find Full Text PDFThe purpose of this work has been to dosimetrically investigate four fluoroscopically guided interventions: the percutaneous vertebroplasty (PVP), the percutaneous disc decompression (PDD), the radiofrequency medial branch neurolysis (RF) (hereafter named spine procedures), and the endovascular treatment for the critical limb ischemia (CLI). The X-ray equipment used was a Philips Integris Allura Xper FD20 imaging system provided with a dose-area product (DAP) meter. The parameters investigated were: maximum skin dose (MSD), air kerma (Ka,r), DAP, and fluoroscopy time (FT).
View Article and Find Full Text PDFBackground: To identify predictive factors of radiation-induced skin toxicity in breast cancer patients by the analysis of dosimetric and clinical factors.
Methods: 339 patients treated between January 2007 and December 2010 are included in the present analysis. Whole breast irradiation was delivered with Conventional Fractionation (CF) (50 Gy, 2.
Purpose: During radiological interventional procedures (RIP) the skin of a patient under examination may undergo a prolonged x-ray exposure, receiving a dose as high as 5 Gy in a single session. This paper describes the use of the OneDose(TM) cable-free system based on p-type MOSFET detectors to determine the entrance skin dose (ESD) at selected points during RIP.
Methods: At first, some dosimetric characteristics of the detector, such as reproducibility, linearity, and fading, have been investigated using a C-arc as a source of radiation.
Background: The OneDosePlusTM system, based on MOSFET solid-state radiation detectors and a handheld dosimetry reader, has been used to evaluate intra-fraction movements of patients with breast and prostate cancer.
Methods: An Action Threshold (AT), defined as the maximum acceptable discrepancy between measured dose and dose calculated with the Treatment Planning System (TPS) (for each field) has been determined from phantom data. To investigate the sensitivity of the system to direction of the patient movements, fixed displacements have been simulated in phantom.
Cone-beam X-ray volumetric imaging in the treatment room, allows online correction of set-up errors and offline assessment of residual set-up errors and organ motion. In this study the registration algorithm of the X-ray volume imaging software (XVI, Elekta, Crawley, United Kingdom), which manages a commercial cone-beam computed tomography (CBCT)-based positioning system, has been tested using a homemade and an anthropomorphic phantom to: (1) assess its performance in detecting known translational and rotational set-up errors and (2) transfer the transformation matrix of its registrations into a commercial treatment planning system (TPS) for offline organ motion analysis. Furthermore, CBCT dose index has been measured for a particular site (prostate: 120 kV, 1028.
View Article and Find Full Text PDF