Publications by authors named "Marc Vila Oliva"
Article Synopsis
- The study proposes a straightforward method for calibrating and validating kilo-voltage cone-beam CT (kV CBCT) models using experimental data.
- Calibration involves a two-step process examining both the source (photon energy spectrum) and the detector (pixel intensity based on photon energy).
- The results show that the method yields high accuracy, with average deviations from experimental data of less than 1.5% for the sources and 4% for energy deposition in detectors across three different kV CBCT systems.
View Article and Find Full Text PDF