Introduction: The purposes of this study were to examine the accuracy and the head positioning effects on measurements of anterior tooth length using 3-dimensional (3D) and conventional dental panoramic radiography and to investigate whether 3D panoramic radiography is suitable for the evaluation of anterior tooth length.
Methods: A simulated human head was radiographed at 4, 8, and 12 mm displaced positions, and at 5°, 10°, and 15° tilted positions from the standard head position using 3D and conventional panoramic radiography, and also using cone-beam computed tomography. Anterior tooth lengths were measured on the panoramic and cone-beam computed tomography images.
Objective: To investigate measurement errors and head positioning effects on radiographs made with new dental panoramic radiograph equipment that uses tomosynthesis.
Materials And Methods: Radiographic images of a simulated human head or phantom were made at standard head positions using the new dental panoramic radiograph equipment. Measurement errors were evaluated by comparing with the true values.