In mammography units, the X-ray tube focal spot is located directly above the chest wall edge of the detector. Therefore, the X-ray beam is incident at a different angle along the cathode-anode axis. When the X-ray beam is incident on the imaging plate (IP) at a relatively large angle, the resolution property of a new imaging plate with dual-sided reading system would be degraded compared with the conventional imaging plate because of the parallax effect, which produces a shift in the image on two sides of the imaging plate.
View Article and Find Full Text PDFPurpose: To investigate the physical imaging properties and detection of simulated microcalcifications of a new computed radiography (CR) system with a pixel size of 50 microm for digital mammography.
Materials And Methods: New and conventional CR were employed in this study. The new CR system included a high-resolution imaging plate coupled with the FCR5000MA (50 microm pixel pitch) including transparent support and a dual-sided reader.