Background And Aims: We assessed the possibility to rule out negative urine cultures by counting with UriSed 3 PRO (77 Elektronika, Hungary) at Helsinki and Uusimaa Hospital District.
Materials And Methods: Bacteria counting of the UriSed 3 PRO automated microscope was verified with reference phase contrast microscopy against growth in culture. After acceptance into routine, results of bacteria and leukocyte counting from 56 426 specimens with eight UriSed 3 PRO instruments were compared against results from parallel samples cultured on chromogenic agar.
Background And Aims: Ten UriSed 3 PRO automated microscopes (77 Elektronika, Hungary) were verified for nine HUSLAB laboratories with 160 000 annual urine samples.
Materials And Methods: Particle counting of the primary UriSed 3 PRO instrument (77 Elektronika, Hungary) was verified against reference visual microscopy with 463 urine specimens, and against urine culture on chromogenic agar plates with parallel 396 specimens. Nine secondary instruments were compared pairwise with the primary instrument.