Publications by authors named "M Pasquino"

Purpose: Aim of the present study is to characterize a deep learning-based auto-segmentation software (DL) for prostate cone beam computed tomography (CBCT) images and to evaluate its applicability in clinical adaptive radiation therapy routine.

Materials And Methods: Ten patients, who received exclusive radiation therapy with definitive intent on the prostate gland and seminal vesicles, were selected. Femoral heads, bladder, rectum, prostate, and seminal vesicles were retrospectively contoured by four different expert radiation oncologists on patients CBCT, acquired during treatment.

View Article and Find Full Text PDF

Objective: To study the effect of beam complexity on VMAT delivery accuracy evaluated by means of a transmission detector, together with the possibility of scoring plan complexity.

Methods: 43 clinical VMAT plans delivered by a TrueBeam linear accelerator to both Delta Discover and Delta Phantom+ for patient-specific quality assurance were evaluated. Global Dose-γ analysis, MLC-γ analysis, percentage of leaves with a deviation between planned and measured leaf tip position lower than 1 mm (LD) were computed.

View Article and Find Full Text PDF

Purpose: To evaluate the dosimetric accuracy of the Delta4 Insight (DI) secondary-check dosimetry system.

Methods: Absolute dosimetry in reference conditions, output factors, percent depth doses normalized and off-axis dose profiles for different field sizes calculated by DI were compared with measurements. Dose calculations for 20 clinical IMRT/VMAT plans generated in the TPS using both AAA or AcurosXB algorithms were compared with measurements.

View Article and Find Full Text PDF

Objective: To investigate the dosimetric behaviour, influence on photon beam fluence and error detection capability of Delta Discover transmission detector.

Methods: The transmission detector (TRD) was characterized on a TrueBeam linear accelerator with 6 MV beams. Linearity, reproducibility and dose rate dependence were investigated.

View Article and Find Full Text PDF

Background: To report 5-year clinical outcomes and toxicity in organ-confined prostate cancer (PCa) for low- and intermediate-risk patients treated with a moderately hypofractionated schedule of radiotherapy (RT) delivered with simultaneous integrated boost (SIB) compared to a conventionally fractionated RT regimen.

Methods: Data of 384 patients with PCa treated between August 2006 and June 2017 were retrospectively reviewed. The treatment schedule consisted of hypofractionated RT (HYPO FR) with SIB up to 70 Gy to the prostate gland and 63 Gy to seminal vesicles delivered in 28 fractions or in conventionally fractionated RT (CONV FR) up to a total dose of 80 Gy in 40 fractions.

View Article and Find Full Text PDF