A high-energy resolution post-column spectrometer for the purpose of electron energy loss spectroscopy (EELS) and energy-filtered TEM in combination with a monochromated (S)TEM is presented. The prism aberrations were corrected up to fourth order using multipole elements improving the electron optical energy resolution and increasing the acceptance of the spectrometer for a combination of object area and collection angles. Electronics supplying the prism, drift tube, high-tension reference and critical lenses have been newly designed such that, in combination with the new electron optics, a sub-50 meV energy resolution has been realized, a 10-fold improvement over past post-column spectrometer designs.
View Article and Find Full Text PDFJ Electron Microsc (Tokyo)
February 2001
A corrected prism-mirror-prism electron energy filter with curved entrance and exit faces was designed and incorporated into a Zeiss EM902 transmission electron microscope. The installation of the new filter required little modification to the existing microscope geometry and electronics. The filter had an energy resolution of 1.
View Article and Find Full Text PDFA novel prism-mirror-prism imaging electron spectrometer with 1 eV energy resolution for a transmission electron microscope permits imaging with spectral energies corresponding to light-optical colour absorptions. The instrument selects the molecular orbital excitations of natural chromophores or of specific dyes normally used in biological light microscopy for delineation and chemical identification, but images them with electron microscopic detail. Heavy atom contrast agents customarily used in electron microscopy are not required.
View Article and Find Full Text PDF