Objective: In implantable cardiac defibrillators (ICDs), long-detection times safely reduce unnecessary and inappropriate therapies. We aimed to evaluate ICD treatment of ventricular arrhythmias in women, compared with men, also taking into account ICD detection.
Methods: The Advance III trial randomised patients implanted with an ICD for primary or secondary prevention in two arms-long and nominal ventricular arrhythmias detection times before therapy delivering (number of intervals needed to detect (NID) 30/40 and 18/24, respectively).
Background: Implantable cardioverter-defibrillator (ICD) therapy has been proven effective in the prevention of sudden cardiac death, but data on outcomes of ICD therapy in the young and otherwise healthy patients with long QT syndrome (LQTS) are limited.
Aim: We sought to collect data on appropriate and inappropriate ICD discharges, risk factors, and ICD-related complications.
Methods: All LQTS patients implanted with an ICD in 14 centres were investigated.
Objectives: This study sought to evaluate the effects of programming a long detection in single-chamber (VVI) implantable cardioverter-defibrillators (ICDs) in the multicenter prospective ADVANCE III (Avoid DeliVering TherApies for Non-sustained Arrhythmias in ICD PatiEnts III) trial.
Background: Programming strategies may reduce unnecessary ICD shocks and their adverse effects but to date have been described only for dual-chamber ICDs.
Methods: A total of 545 subjects (85% male; atrial fibrillation 25%, left ventricular ejection fraction 31%, ischemic etiology 68%, secondary prevention indications 32%) receiving a VVI ICD were randomized to long detection (30 of 40 intervals) or standard programming (18 of 24 intervals) based on device type, atrial fibrillation history, and indication.
Aims: A long-detection interval (LDI) (30/40 intervals) has been proved to be superior to a standard-detection interval (SDI) (18/24 intervals) in terms of reducing unnecessary implantable cardioverter defibrillator (ICD) therapies. To better evaluate the different impact of LDI and anti-tachycardia pacing (ATP) on reducing painful shocks, we assessed all treated episodes in the ADVANCE III trial.
Methods And Results: A total of 452 fast (200 ms < cycle length ≤ 320 ms) arrhythmic episodes were recorded: 284 in 138 patients in the SDI arm and 168 in 82 patients in the LDI arm (106/452 inappropriate detections).