Objective: The aim of our study was (1) to establish an in-bore targeted biopsy of suspicious prostate lesions, avoiding bowel penetration using a transgluteal approach and (2) to assess operator setup, patient comfort and safety aspects in the clinical setting for freehand real-time MR-guidance established for percutaneous procedures in an open MR-scanner.
Material And Methods: 30 patients with suspect prostate lesions were biopsied in a cylindrical 3T-MRI system using a transgluteal approach in freehand technique. One to three biopsies were sampled using continuous dynamic imaging.