Publications by authors named "Lorenzo Radici"

Purpose: Aim of the present study is to characterize a deep learning-based auto-segmentation software (DL) for prostate cone beam computed tomography (CBCT) images and to evaluate its applicability in clinical adaptive radiation therapy routine.

Materials And Methods: Ten patients, who received exclusive radiation therapy with definitive intent on the prostate gland and seminal vesicles, were selected. Femoral heads, bladder, rectum, prostate, and seminal vesicles were retrospectively contoured by four different expert radiation oncologists on patients CBCT, acquired during treatment.

View Article and Find Full Text PDF

Objective: To study the effect of beam complexity on VMAT delivery accuracy evaluated by means of a transmission detector, together with the possibility of scoring plan complexity.

Methods: 43 clinical VMAT plans delivered by a TrueBeam linear accelerator to both Delta Discover and Delta Phantom+ for patient-specific quality assurance were evaluated. Global Dose-γ analysis, MLC-γ analysis, percentage of leaves with a deviation between planned and measured leaf tip position lower than 1 mm (LD) were computed.

View Article and Find Full Text PDF

Purpose: To evaluate the dosimetric accuracy of the Delta4 Insight (DI) secondary-check dosimetry system.

Methods: Absolute dosimetry in reference conditions, output factors, percent depth doses normalized and off-axis dose profiles for different field sizes calculated by DI were compared with measurements. Dose calculations for 20 clinical IMRT/VMAT plans generated in the TPS using both AAA or AcurosXB algorithms were compared with measurements.

View Article and Find Full Text PDF

Proper delineation of both target volumes and organs at risk is a crucial step in the radiation therapy workflow. This process is normally carried out manually by medical doctors, hence demanding timewise. To improve efficiency, auto-contouring methods have been proposed.

View Article and Find Full Text PDF

We previously reported on a cohort of breast cancer patients affected with ductal carcinoma in situ (DCIS) that were treated with breast conservative surgery and hypofractionated whole-breast radiotherapy with a concomitant boost to the lumpectomy cavity. We now report on the long-term results of the oncological and toxicity outcomes, at a median follow-up of 11.2 years.

View Article and Find Full Text PDF

The optimal timing of surgery after neoadjuvant chemoradiotherapy (CRT) in locally advanced rectal cancer is still controversial. Aim of this study was to evaluate the effect of increasing time interval between the end of CRT and surgery on pathological outcomes. This is a retrospective analysis on 114 patients treated with long-course neoadjuvant RT with or without chemotherapy between January 2005 and September 2020.

View Article and Find Full Text PDF

Objective: To investigate the dosimetric behaviour, influence on photon beam fluence and error detection capability of Delta Discover transmission detector.

Methods: The transmission detector (TRD) was characterized on a TrueBeam linear accelerator with 6 MV beams. Linearity, reproducibility and dose rate dependence were investigated.

View Article and Find Full Text PDF

Purpose: To investigate the performance of various algorithms for deformable image registration (DIR) for propagating regions of interest (ROIs) using multiple commercial platforms, from computed tomography to cone beam computed tomography (CBCT) and megavoltage computed tomography.

Methods And Materials: Fourteen institutions participated in the study using 5 commercial platforms: RayStation (RaySearch Laboratories, Stockholm, Sweden), MIM (Cleveland, OH), VelocityAI and SmartAdapt (Varian Medical Systems, Palo Alto, CA), and ABAS (Elekta AB, Stockholm, Sweden). Algorithms were tested on synthetic images generated with the ImSimQA (Oncology Systems Limited, Shrewsbury, UK) package by applying 2 specific deformation vector fields (DVF) to real head and neck patient datasets.

View Article and Find Full Text PDF

Objective: The aim of this work was to investigate the main dosimetric characteristics and the performance of an A26 Exradin ionization microchamber (A26 IC) and a W1 Exradin plastic scintillation detector (W1 PSD) in small photon beam dosimetry for treatment planning system commissioning and quality assurance programme.

Methods: Detector characterization measurements (short-term stability, dose linearity, angular dependence and energy dependence) were performed in water for field sizes up to 10 × 10 cm. Polarity effect (P) was examined for the A26 IC.

View Article and Find Full Text PDF