Med Ultrason
February 2021
Aims: To determine the reliability of transpalpebral ultrasound in B-mode (B-TUS) with a high-resolution linear probe (18 MHz) in estimating both the ocular anterior chamber depth (ACD) and axial length (AL), as well as its agreement with the IOL Master 500 optical biometer.
Material And Methods: Cross-sectional study on 82 eyes of 41 volunteers with no history of eye disease. ACD and AL were determined using B-TUS and the IOL Master 500.
Purpose: To evaluate the reliability and agreement between Fully Refocused Steady-State magnetic resonance sequences (FRSS) and the IOLMaster 500 optical biometer for measuring anterior chamber depth (ACD) and axial length (AL).
Methods: In a sample of 32 healthy volunteers, separate observers measured the ACD and AL of both eyes using both techniques (inter-method) and through repeated FRSS measurements (interobserver) and by the same observer (intraobserver). We employed the Bland-Altman method to determine the agreement between FRSS and partial coherence interferometry (using the IOLMaster) and the interobserver and intraobserver variability, providing the limits of agreement (LoA, or mean difference ± 1.