Background: A frequent problem in dental units is the microbial contamination of water and biofilm formation in the water supply lines. After random identification of a bacterial contaminated dental unit (310 cfu/ml) in a practise with 3 dental units we implemented the present study to evaluate the efficacy of the PotoClean(®) technology, based on anodic oxidation.
Method: The efficacy of a regular low concentrated permanent decontamination (1 mg Cl/L) with an additional intensive decontamination by PotoClean(®) (three times 20 mg Cl/ml for 2 h) on three dental units was tested over 7 months.