Background: Independent testing of home blood pressure (BP) measurement (HBPM) devices is often lacking, particularly among older and multi-morbid patients.
Methods: We studied the Bpro G2 (using tonometry), Omron HeartGuide (using occlusive oscillometric technology), and Heartisans (using photoplethysmography) wrist watch HBPM devices against a gold standard brachial sphygmomanometer. To test device performance, we used the ISO81060-2 protocol (though this protocol cannot formally validate cuffless devices).