A comprehensive comparative validation for two different types of dissolved oxygen (DO) analyzers, amperometric and optical, is presented on two representative commercial DO analyzers. A number of performance characteristics were evaluated including drift, intermediate precision, accuracy of temperature compensation, accuracy of reading (under different measurement conditions), linearity, flow dependence of the reading, repeatability (reading stability), and matrix effects of dissolved salts. The matrix effects on readings in real samples were evaluated by analyzing the dependence of the reading on salt concentration (at saturation concentration of DO).
View Article and Find Full Text PDFThis work introduces a conceptually new approach of measuring pH of mixed-solvent liquid chromatography (LC) mobile phases. Mobile phase pH is very important in LC, but its correct measurement is not straightforward, and all commonly used approaches have deficiencies. The new approach is based on the recently introduced unified pH (pH(abs)) scale, which enables direct comparison of acidities of solutions made in different solvents based on chemical potential of the proton in the solutions.
View Article and Find Full Text PDFAnal Chim Acta
September 2012
A high-accuracy Winkler titration method has been developed for determination of dissolved oxygen concentration. Careful analysis of uncertainty sources relevant to the Winkler method was carried out and the method was optimized for minimizing all uncertainty sources as far as practical. The most important improvements were: gravimetric measurement of all solutions, pre-titration to minimize the effect of iodine volatilization, accurate amperometric end point detection and careful accounting for dissolved oxygen in the reagents.
View Article and Find Full Text PDFSensors (Basel)
December 2013
This tutorial focuses on measurement uncertainty estimation in amperometric sensors (both for liquid and gas-phase measurements). The main uncertainty sources are reviewed and their contributions are discussed with relation to the principles of operation of the sensors, measurement conditions and properties of the measured samples. The discussion is illustrated by case studies based on the two major approaches for uncertainty evaluation-the ISO GUM modeling approach and the Nordtest approach.
View Article and Find Full Text PDFIn this report a gravimetric micro-Winkler titration method for determination of dissolved oxygen concentration in water is presented. Mathematical model of the method taking into account all influence factors is derived and an uncertainty analysis is carried out to determine the uncertainty contributions of all influence factors. The method is highly accurate: the relative expanded uncertainties (k=2) are around 1% in the case of small (9-10 g) water samples.
View Article and Find Full Text PDF