Publications by authors named "Lauren A Hill"

We present a new chemiluminescence detector, with solution channels that have been machined into a Teflon disk and sealed with a sapphire window. The configuration of the flow cell can be conveniently modified by replacing the Teflon disk. A comparison of some existing and novel designs, using the chemiluminescence reaction of morphine with acidic potassium permanganate and the bioluminescence reaction of ATP with the commercially available "BacTiter-Glo" reagent, has revealed that a serpentine channel allows greater quantities of light to be captured than a spiral channel, due to more efficient mixing of the analyte and reagent solutions within the cell.

View Article and Find Full Text PDF

A sequential injection analysis procedure with dual-reagent chemiluminescence detection was applied to the screening of street drug seizure samples for the presence of heroin. The chemiluminescence reagents (acidic potassium permanganate and tris(2,2'-bipyridine)ruthenium(III)) were aspirated from either side of a sample aliquot that was sufficiently large to prevent interdispersion of the reagent zones, and therefore two different chemical reactions could be performed simultaneously at either end of the sample zone. The presence of heroin in seizure samples was indicated by a strong response with the tris(2,2'-bipyridine)ruthenium(III) reagent and confirmed by a significant increase in the response with the permanganate reagent when the sample was treated with sodium hydroxide to hydrolyse the heroin to morphine.

View Article and Find Full Text PDF