Objectives: This study sought to evaluate the effects of programming a long detection in single-chamber (VVI) implantable cardioverter-defibrillators (ICDs) in the multicenter prospective ADVANCE III (Avoid DeliVering TherApies for Non-sustained Arrhythmias in ICD PatiEnts III) trial.

Background: Programming strategies may reduce unnecessary ICD shocks and their adverse effects but to date have been described only for dual-chamber ICDs.

Methods: A total of 545 subjects (85% male; atrial fibrillation 25%, left ventricular ejection fraction 31%, ischemic etiology 68%, secondary prevention indications 32%) receiving a VVI ICD were randomized to long detection (30 of 40 intervals) or standard programming (18 of 24 intervals) based on device type, atrial fibrillation history, and indication.

View Article and Find Full Text PDF

Aims: We sought to determine whether atrioventricular junction ablation (AVJA) in patients with cardiac resynchronization therapy (CRT) implantable cardioverter-defibrillator (ICD) and with permanent atrial fibrillation (AF) has a positive impact on ICD shocks and hospitalizations compared with rate-slowing drugs.

Methods And Results: This is a pooled analysis of data from 179 international centres participating in two randomized trials and one prospective observational research. The co-primary endpoints were all-cause ICD shocks and all-cause hospitalizations.

View Article and Find Full Text PDF

Aims: A long-detection interval (LDI) (30/40 intervals) has been proved to be superior to a standard-detection interval (SDI) (18/24 intervals) in terms of reducing unnecessary implantable cardioverter defibrillator (ICD) therapies. To better evaluate the different impact of LDI and anti-tachycardia pacing (ATP) on reducing painful shocks, we assessed all treated episodes in the ADVANCE III trial.

Methods And Results: A total of 452 fast (200 ms < cycle length ≤ 320 ms) arrhythmic episodes were recorded: 284 in 138 patients in the SDI arm and 168 in 82 patients in the LDI arm (106/452 inappropriate detections).

View Article and Find Full Text PDF

Background: Three trials demonstrated recently that a long detection window reduces implantable cardioverter-defibrillator (ICD) therapy in primary prevention patients. Avoid Delivering Therapies for Nonsustained Arrhythmias in ICD Patients III (ADVANCE III) was the only trial that enrolled both primary and secondary prevention patients.

Methods And Results: Of the 1902 patients enrolled in the ADVANCE III trial, 477 received a defibrillator for secondary prevention; 248 patients were randomly assigned to a long detection setting (30 of 40 intervals) and 229 to the nominal setting (18 of 24 intervals) for ventricular arrhythmias with cycle length ≤ 320 ms.

View Article and Find Full Text PDF

Importance: Using more intervals to detect ventricular tachyarrhythmias has been associated with reducing unnecessary implantable cardioverter-defibrillator (ICD) therapies.

Objective: To determine whether using 30 of 40 intervals to detect ventricular arrhythmias (VT) (long detection) during spontaneous fast VT episodes reduces antitachycardia pacing (ATP) and shock delivery more than 18 of 24 intervals (standard detection).

Design, Setting, And Participants: Randomized, single-blind, parallel-group trial that enrolled 1902 primary and secondary prevention patients (mean [SD] age, 65 [11] years; 84% men; 75% primary prevention ICD) with ischemic and nonischemic etiology undergoing first ICD implant at 1 of 94 international centers (March 2008-December 2010).