Background: Electromagnetic Tracking (EMT) technology has been integrated in a prototype high-dose-rate brachytherapy (HDR-BT) afterloading device. Its potential for dwell position (DP) monitoring has earlier been demonstrated in prostate phantoms. However, its performance for prostate BT in the clinical setting remains to be assessed.
View Article and Find Full Text PDFPurpose: Deformable image registration (DIR) can be used to accumulate the absorbed dose distribution of daily image-guided adaptive external beam radiation treatment (EBRT) and brachytherapy (BT). Since dose-volume parameter addition assumes a uniform delivered EBRT dose around the planned BT boost, the added value of DIR over direct addition was investigated for dose accumulation in bladder and rectum.
Material And Methods: For 10 patients (EBRT 46/46.
Int J Radiat Oncol Biol Phys
November 2017
Purpose: To quantify magnetic resonance imaging (MRI) distortions on a plastic intracavitary/interstitial applicator with plastic needles at a field strength of 3 T and to determine the dosimetric impact, using patient data.
Methods And Materials: For 11 cervical cancer patients, our clinical MRI protocol was extended with 3 scans. From the first scan, a multi-echo acquisition, a map of the magnetic field (B) was calculated and used to quantify the field inhomogeneity.
Purpose: Structure-based deformable image registration (DIR) can be used to calculate accumulated dose volume histogram parameters for cervical cancer brachytherapy (BT). The purpose of this study is to investigate dose warping uncertainties for the accumulated dose to the 2 cm receiving the highest dose [Formula: see text] in the rectal wall, using a physically realistic model (PRM) describing rectal wall deformation.
Methods And Materials: For 10 patients, treated with MRI-guided pulsed dose rate BT (two times 24 × 0.
Background And Purpose: Structure-based deformable image registration (DIR) can be used to calculate accumulated brachytherapy (BT) and external-beam radiation therapy (EBRT) dose-volume histogram (DVH) parameters in cervical cancer. Since direct parameter addition does not take dose non-uniformity into account, the added value of DIR over addition methods was investigated for bladder and rectum.
Materials And Methods: For twelve patients (EBRT: 46Gy, EBRT+BT: D 85-90Gy in equivalent dose in 2Gy fractions) the EBRT planning CT and BT planning MRI were registered using DIR.