Article Synopsis
  • The paper evaluates the performance of the Kodak EKTACHEM GLU/BUN Analyzer for measuring glucose and urea using established NCCLS protocols.
  • It involved a multicentre trial across four European countries and employed specific experiments to assess precision and accuracy.
  • The study highlighted challenges in compliance and sample collection for comparative measurements while noting consistent results across different laboratories in replication tests.
View Article and Find Full Text PDF

The study carried out consists of evaluating the results obtained for estimation of glucose and urea with a procedure the originality of which resides mainly in the use of preconditioned reagents, in the dry form, laid out on a support where the colour develops, proceeding to a reading by reflectometry and linearise the response by a procedure of calculation carried out by a microprocessor. The evaluation of the results obtained with the Kodak-Ektachem prototype for estimations of glucose and urea in the plasma was carried out according to the NCCLS protocol in the field of a coordinated European study. The imprecision was estimated at several levels of concentration after a study of intraserial and day to day reproducibility with and without contamination correction, the variation coefficients obtained from day to day are less than 3% when the concentrations of glucose and urea are higher than 5 mmol.