Transvalvular turbulence caused by the blood flow in implanted cardiac valvular prosthesis is the cause of thrombosis at the post-valvular region. The literature in this connection indicates that thromboembolic complications are more common with the tilting disc mechanical valves. In order to study the turbulence caused by porcine biological prosthetic valves, by a caged-ball Biomed mechanical valve and by a tilting disc Björk-Shiley valve, a device was developed which permits to visualize and to photograph flow characteristics in the three types of valves with the purpose of determining the degree of turbulence generated in each case.