Purpose: Experimental patient-specific QA (PSQA) is a time and resource-intensive process, with a poor sensitivity in detecting errors. Radiation therapy facilities aim to substitute it by means of independent dose calculation (IDC) in combination with a comprehensive beam delivery QA program. This paper reports on the commissioning of the IDC software tool myQA iON (IBA Dosimetry) for proton therapy and its clinical implementation at the MedAustron Ion Therapy Center.

View Article and Find Full Text PDF

Purpose: To fully characterize the flat panel detector of the new Sphinx Compact device with scanned proton and carbon ion beams.

Materials And Methods: The Sphinx Compact is designed for daily QA in particle therapy. We tested its repeatability and dose rate dependence as well as its proportionality with an increasing number of particles and potential quenching effect.

View Article and Find Full Text PDF
Article Synopsis
  • The study reports on the first use of the IBA Sphinx/Lynx device for daily Quality Assurance (QA) in proton and carbon ion therapies.
  • Over 3 years of data collected for protons and 2 years for carbon ions focused on key QA parameters like spot size, beam range, and dose.
  • The results indicate that the device effectively detects performance issues and establishes actionable tolerances, enhancing the QA process in a dual particle therapy facility.
View Article and Find Full Text PDF

This paper reviews the ecosystem of GATE, an open-source Monte Carlo toolkit for medical physics. Based on the shoulders of Geant4, the principal modules (geometry, physics, scorers) are described with brief descriptions of some key concepts (Volume, Actors, Digitizer). The main source code repositories are detailed together with the automated compilation and tests processes (Continuous Integration).

View Article and Find Full Text PDF

Purpose: The Local Effect Model version one (LEM I) is applied clinically across Europe to quantify the relative biological effectiveness (RBE) of carbon ion beams. It requires the full particle fluence spectrum differential in energy in each voxel as input parameter. Treatment planning systems (TPSs) use beamline-specific look-up tables generated with Monte Carlo (MC) codes.