Publications by authors named "L A Whitford"

This study evaluated a range of programming parameters available for the ACE and CIS speech-processing strategies in the Nucleus 24 cochlear implant system. Specifically, the effect on speech perception of adjustments to the number of channels and the stimulation rate in the CIS strategy, and the effect of adjustments to the number of maxima in the ACE strategy were investigated in a group of adult subjects. Based on these findings, and the results of a previous study (Vandali et al.

View Article and Find Full Text PDF

This study measured changes in electrode impedance over time in 19 adults and 29 children implanted with the Nucleus 24 cochlear implant system, using common ground and three monopolar modes of stimulation, over a series of time intervals. Impedances increased from the intraoperative to the initial fitting session, decreased during the initial fitting session and for the following week, and were then stable. Impedances were lowest for the common ground mode, while for the monopolar modes, impedances were related to the surface area of the return electrode.

View Article and Find Full Text PDF

Objective: The effect on speech perception of using higher stimulation rates than the 14.4 kHz available in the Nucleus 24 cochlear implant system was investigated. The study used the Nucleus Research Platform 8 (RP8) system, comprising the CI24RE receiver-stimulator with the Contour electrode array, the L34SP body-worn research speech processor, and the Nucleus Programming Environment (NPE) fitting and Neural Response Telemetry (NRT) software.

View Article and Find Full Text PDF

Objective: The objective of this study was to evaluate streamlined programming procedures for the Nucleus cochlear implant system with the Contour electrode array.

Design: Phase 1 involved an examination of the clinical MAPs for the first 103 recipients implanted with the Contour electrode array in the Melbourne Cochlear Implant Clinic, to examine the ability to predict the entire MAP based on a smaller number of clinically determined T- and/or C-levels. In phase 2, a subset of the streamlined procedures was selected and clinically evaluated, using speech perception and subjective preference measures.